Pipette Calibration Services UK
Our Services
At AccuCal, we offer a range of pipette calibration services designed to meet the diverse needs of our clients. Our services are tailored to ensure your pipettes perform with the highest precision, whether you are using single-channel or multi-channel pipettes.
Explore our service options below:
Single-Channel Pipette Calibration
Volumes: 1 volume
Readings: 4 readings
Volumes: 2 volumes
Readings: 4 readings each
Volumes: 3 volumes
Readings: 3 readings each
Volumes: 3 volumes
Readings: 5 readings each
Multi-Channel Pipette Calibration
Volumes: 1 volume
Readings: 4 readings
Volumes: 2 volumes
Readings: 2 readings each
Volumes: 3 volumes
Readings: 2 readings each
Volumes: 3 volumes
Readings: 3 readings each
Customised Test Plans
Every laboratory has unique requirements. That is why we offer the flexibility for you to customise your own calibration test plan — so the service matches your exact application, quality system, or compliance needs.
What You Can Customise
- Specific volumes tested across the pipette's working range
- Number of readings per volume (from 4 up to 10 or more)
- Tolerance limits — ISO 8655, manufacturer specifications, or your own customer-defined values
- Reporting format to match your existing quality system
Standards and Compliance
Our calibration methodology follows ISO 8655 guidelines, ensuring measurements are performed to a recognised international standard whichever test plan you choose.
Customised test plans are available with the ISO 8655 or Customer specification options.
Our Post Service and Calibration Process
Every pipette we receive goes through three clearly defined stages — so you know exactly what you're paying for and exactly what's been done.
Step 1 — Servicing
We disassemble each pipette and thoroughly clean the internal parts, including the piston, applying manufacturer-suggested lubricant. We check the performance of the O-ring and seal. Spare parts are not included as standard — if a part replacement is needed, we will always seek your authorisation before fitting it, so there are no surprise costs.
Step 2 — Calibration
We offer multiple levels of calibration so you can choose the depth of service that matches your application. Our calibrations range from a quick two-volume check with four readings per volume — ideal for routine verification — up to a three-volume check with ten readings per volume for higher confidence and more detailed documentation.
All measurements are performed on calibrated analytical balances in a temperature- and humidity-controlled environment, and checked against manufacturer specifications, the ISO 8655 standard, or your own customer-defined tolerances.
Step 3 — Finalisation
Once the pipette passes calibration, we clean the external body with isopropyl alcohol and apply an electronic identification sticker showing: calibration date, next due date, serial number, certificate number, engineer's initials, and a scannable QR code linking back to the certificate.
Pre-Calibration Service (As Found)
Our pre-calibration (As Found) service assesses the current performance of your pipette without any servicing or adjustments. Ideal for routine verification, baseline checks before further work, or simply confirming a pipette is operating within specification.
Step 1 — Performance Check
We calibrate your pipette in its current state, without prior cleaning, lubrication, or part replacement. Calibration is performed at the volumes and reading depth appropriate to the level you choose — from a quick two-volume check with four readings per volume, up to a three-volume check with ten readings per volume. All measurements are taken on calibrated analytical balances in a temperature- and humidity-controlled environment, and compared against manufacturer specifications, the ISO 8655 standard, or your own customer-defined tolerances.
Step 2 — Certification and Reporting
Each pipette receives a Certificate of Calibration indicating whether it has passed or failed against the chosen specification, along with the measured values, deviation from nominal, standard deviation, and tolerance limits applied. An electronic identification sticker is also applied to the pipette showing the next calibration due date, serial number, certificate number, engineer's initials, and a scannable QR code linking back to the certificate.